Home      Products     Endress+Hauser     Analysis       CPS91E Digital pH sensor
  • CPS91E Digital pH sensor
  • CPS91E Digital pH sensor
CPS91E Digital pH sensorCPS91E Digital pH sensor

CPS91E Digital pH sensor

  • Product Item : CPS91E-AA7BTH2
  • Category: Analysis
  • Measuring principle: Potentiometric
  • Application: Process technology and monitoring of processes with: • Rapidly changing pH values • High proportion of electrode poisons such as H2S
  • Characteristic: Digital pH electrode for chemical process with an ion trap for poison-resistant reference
  • Measuring principle: Potentiometric Application Process technology and monitoring of processes with: • Rapidly changing pH values • High proportion of electrode poisons such as H2S Characteristic D
  • INQUIRY   EMAIL
无标题文档

Digital pH sensor Memosens CPS91E

Memosens 2.0 pH electrode for chemical processes and poisoning media

Field of application

Memosens CPS71E is specially designed for demanding processes. Its unique, contamination-resistant reference and its resistance to moisture ensure reliable measurement even in heavily polluted, aggressive media such as strong acids and bases. Thanks to Memosens 2.0 digital technology, CPS71E offers extended storage of calibration and process data providing the perfect basis for predictive maintenance. Pre-calibration in the lab and quick sensor exchange on site maximize process uptime.

Processes in chemical industry and monitoring of processes with:

  • High proportion of electrode poisons such as H2S

  • Quickly changing pH values

With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.

Benefits

  • The electrode's unique ion trap prevents poisoning of the junction and reference, making it resistant to strong acids and bases and ensuring a long sensor lifetime.

  • The optional pressurized reference allows for reliable measurement in blocking media such as dispersions.

  • Flexible installation thanks to optional upside-down mounting

  • Inductive cable connection and non-contact signal transmission eliminate any problems due to moisture or corrosion and increase process integrity.

  • Fast sensor exchange on site reduces process downtime and operating costs.

  • Memosens 2.0 digital technology makes trend identification and predictive maintenance possible thanks to its extended storage of calibration and process data. The technology paves the way for advanced IIoT services.

  • Measuring principle

    Potentiometric

  • Application

    Process technology and monitoring of processes with:
    • Rapidly changing pH values
    • High proportion of electrode poisons such as H2S

  • Characteristic

    Digital pH electrode for chemical process with an ion trap for poison-resistant reference

  • Measurement range

    Application B:
    • pH: 0 to 14
    Application H:
    • pH: 0 to 12

  • Measuring principle

    Gel compact electrode with ceramic junction and ion trap

  • Design

    All shaft lengths with temperature sensor
    Advanced gel technology

  • Material

    Sensor shaft: Glass to suit process
    pH membrane glass: Type B, Type N
    Metal lead: Ag/AgCl
    Open aperture: Ceramic junction, zirconium dioxide
    O-ring: FKM
    Process coupling: PPS fiber-glass reinforced
    Nameplate: Ceramic metal oxide

  • Dimension

    Diameter: 12 mm (0.47 inch)
    Shaft lengths: 120, 225, 360 and 425 mm
    (4.72, 8.86, 14.2 and 16.7 inch)

  • Process temperature

    1 to 140 °C (32 to 284 °F)

  • Process pressure

    0,8 to 14 bar (11,6 to 203 psi) (absolute)

  • Temperature sensor

    NTC 30k

  • Ex certification

    With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
    hazardous areas Zone 0, Zone 1 and Zone 2.

  • Connection

    Inductive, contactless connection head with Memosens 2.0 technology

  • Ingress protection

    IP68

  • Additional certifications

    Additional certifications

CPS91D-7BT21 replaced by CPS91E-AA7BTH2

Basic model
Orbipore CPS91D Memosens 
Version:
7 basic version 
Application Range:
BT 0-14pH; 0-110oC; 13 bar; with ion trap 
Shaft Length:
2 120mm 
Approval:
1 Non-hazardous area 




CPS91E-AA7BTH2


  • Approval
    AA
    Non-hazardous area
  • Electrode Type
    7
    Basic version, zero point pH 7,0,temperature sensor NTC 30k
  • Application Range
    B
    0-14 pH, 0...110oC, 0,8...14 bar (abs)
  • Reference System
    TH
    Open junction, ion trap,3 M KCl, Ag/AgCl
  • Shaft Length
    2
    120mm


CPS91D-7BO21 replaced by CPS91E-AA7BHH2

Basic model
Orbipore CPS91D Memosens 
Version:
7 basic version 
Application Range:
BO 0-14pH; 0-110oC. 
Shaft Length:
2 120mm 
Approval:
1 Non-hazardous area 

CPS91E-AA7BHH2


  • Approval
    AA
    Non-hazardous area
  • Electrode Type
    7
    Basic version, zero point pH 7,0,temperature sensor NTC 30k
  • Application Range
    B
    0-14 pH, 0...110oC, 0,8...14 bar (abs)
  • Reference System
    HH
    Open junction, 3 M KCl, Ag/AgCl
  • Shaft Length
    2
    120mm