Continuous / Liquids pressure measurement
|
Measuring Principle
|
Differential pressure
|
Characteristic / Application
|
Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
|
Supply / Communication
|
4...20mA HART:
11,5...45V DC
Ex ia: 11,5...30V DC
|
Accuracy
|
0,1%
"PLATINUM" 0,075%
|
Long term stability
|
0,05% of URL/year
0,125% of URL/5 years
|
Ambient temperature
|
-40°C...85°C
(-40°F...185°F)
|
Process temperature
|
-40°C...120°C
(-40°F...248°F)
|
Process pressure absolute / max. overpressure limit
|
-40°C...120°C
(-40°F...248°F)
|
Pressure measuring range
|
10mbar...40bar
(0.15...580psi)
|
Main wetted parts
|
316L
|
Process connection
|
1/4-18 NPT
|
Output
|
4...20mA HART
PROFIBUS PA
FOUNDATION Fieldbus
|
Certificates / Approvals
|
ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI
SIL
NACE
EN10204-3.1
|
Options
|
4-line digital display
Aluminium housing
|
Application limits
|
Measuring cell:
Metal welded
|
|
Gas Pressure Measurement
|
Measuring Principle
|
Differential pressure
|
Product headline
|
Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
|
Max. measured error
|
0,1%
"PLATINUM" 0,075%
|
Max. process pressure
|
10mbar...40bar
(0.15...580psi)
|
Medium temperature range
|
Temperature gradient from pressure piping
|
Display/Operation
|
Option
|
Outputs
|
4...20mA HART
|
Digital communication
|
HART
|
Hazardous area approvals
|
ATEX
FM
CSA
IECEx
|
|
Liquids pressure measurement
|
Measuring Principle
|
Differential pressure
|
Product headline
|
Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
|
Max. measured error
|
0,1%
"PLATINUM" 0,075%
|
Max. process pressure
|
10mbar...40bar
(0.15...580psi)
|
Medium temperature range
|
Temperature gradient from pressure piping
|
Display/Operation
|
Option
|
Outputs
|
4...20mA HART
|
Digital communication
|
HART
|
Hazardous area approvals
|
ATEX
FM
CSA
IECEx
|
|
Pressure measurement
|
Measuring Principle
|
Differential pressure
|
Characteristic
|
Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
|
Supply voltage
|
4...20mA HART
11,5...45V DC (Non Ex):
Ex ia: 11,5...30V DC
PROFIBUS PA:
9...32 V DC (Non Ex)
FOUNDATION Fieldbus:
9...32 V DC (Non Ex)
|
Reference Accuracy
|
0,1%
"PLATINUM" 0,075%
|
Long term stability
|
0,05% of URL/year
0,125% of URL/5 years
|
Process temperature
|
-40°C...120°C
(-40°F...248°F)
|
Ambient temperature
|
-40°C...85°C
(-40°F...185°F)
|
Measuring cell
|
10mbar...40bar
(0.15...580psi)
|
Smallest calibratable span
|
0.5 mbar (0.0075 psi)
|
Max. Turn down
|
100:1
|
Max. overpressure limit
|
on one side:
160 bar
(2300 psi)
|
Process connection
|
1/4-18 NPT
|
Material process membrane
|
316L, AlloyC,
|
Material gasket
|
Viton, PTFE, EPDM, NBR
|
Fill fluid
|
Silicone oil,
Inert oil,
|
Material housing
|
Die-cast aluminum
|
Output
|
4...20mA HART
PROFIBUS PA
FOUNDATION Fieldbus
|
Certificates / Approvals
|
ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI
NACE
SIL
EN10204-3.1
|
Options
|
4-line digital display
Aluminium housing
|
|
Steam Gas
|
Measuring Principle
|
Differential pressure
|
Product headline
|
Digital transmitter with metallic measuring diaphragms
Compact size
Modular transmitter
Long-term stability
|
Max. measured error
|
0,1%
"PLATINUM" 0,075%
|
Max. process pressure
|
10mbar...40bar
(0.15...580psi)
|
Medium temperature range
|
Temperature gradient from pressure piping
|
Display/Operation
|
Option
|
Outputs
|
4...20mA HART
|
Digital communication
|
HART
|
Hazardous area approvals
|
ATEX
FM
CSA
IECEx
|
|