Gas
|
Measuring Principle
|
Vortex
|
Product headline
|
The flowmeter for accurate steam flow measurement with wet steam detection, available as compact or remote version. Integrated temp. measuring for mass/ energy flow of saturated steam. Suitable for a wide range of applications; optimized for steam applications.
|
Sensor features
|
Highest process safety – dualsens version enables redundant measurement. High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer. No maintenance – lifetime calibration. Wet steam detection and measurement for DN 25 to 100 (1 to 4"). Inlet run compensation.
|
Transmitter features
|
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology™. Display module with data transfer function. Robust two-chamber housing.
|
Nominal diameter range
|
DN 15 to 300 (½ to 12")
|
Wetted materials
|
Measuring tube: 1.4408 (C3FM); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4435 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Connection: 1.4404 (F316/F316L); CX2MW similar to Alloy C22, 2.4602; 1.4408 (CF3M)
|
Measured variables
|
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
|
Max. measured error
|
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (liquid): ±0.85%
Mass flow (steam, gas): ±1.7 %
|
Measuring range
|
Liquid: 0.16 to 2412 m³/h (0.09 to 1420 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 2 to 32 166 m³/h (1.18 to 18 932 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
|
Max. process pressure
|
PN 40, Class 300, 20K
|
Medium temperature range
|
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)
|
Ambient temperature range
|
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
|
Sensor housing material
|
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
|
Transmitter housing material
|
AlSi10Mg, coated; 1.4404 (316L)
|
Degree of protection
|
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
|
Display/Operation
|
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
|
Outputs
|
4-20 mA HART (passive)
4-20 mA (passive)
Pulse/frequency/switch output (passive)
|
Inputs
|
4-20 mA (passive)
|
Digital communication
|
HART, PROFIBUS PA, FOUNDATION Fieldbus
|
Power supply
|
DC 12 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4-20 mA HART, 4-20 mA)
DC 12 to 35 V (4-20 mA HART, pulse/frequency/switch output, 4-20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
|
Hazardous area approvals
|
ATEX, IECEx, cCSAus
|
Other approvals and certificates
|
3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, functional safety
PED, CRN, AD 2000
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614-1, similar to ASME IX (on request)
|
|
Liquids
|
Measuring Principle
|
Vortex
|
Product headline
|
The flowmeter for accurate steam flow measurement with wet steam detection, available as compact or remote version. Integrated temp. measuring for mass/ energy flow of saturated steam. Suitable for a wide range of applications; optimized for steam applications.
|
Sensor features
|
Highest process safety – dualsens version enables redundant measurement. High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer. No maintenance – lifetime calibration. Wet steam detection and measurement for DN 25 to 100 (1 to 4"). Inlet run compensation.
|
Transmitter features
|
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology™. Display module with data transfer function. Robust two-chamber housing.
|
Nominal diameter range
|
DN 15 to 300 (½ to 12")
|
Wetted materials
|
Measuring tube: 1.4408 (C3FM); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4435 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Connection: 1.4404 (F316/F316L); CX2MW similar to Alloy C22, 2.4602; 1.4408 (CF3M)
|
Measured variables
|
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
|
Max. measured error
|
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (liquid): ±0.85%
Mass flow (steam, gas): ±1.7 %
|
Measuring range
|
Liquid: 0.16 to 2412 m³/h (0.09 to 1420 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 2 to 32 166 m³/h (1.18 to 18 932 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
|
Max. process pressure
|
PN 40, Class 300, 20K
|
Medium temperature range
|
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)
|
Ambient temperature range
|
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
|
Sensor housing material
|
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
|
Transmitter housing material
|
AlSi10Mg, coated; 1.4404 (316L)
|
Degree of protection
|
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
|
Display/Operation
|
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
|
Outputs
|
4-20 mA HART (passive)
4-20 mA (passive)
Pulse/frequency/switch output (passive)
|
Inputs
|
4-20 mA (passive)
|
Digital communication
|
HART, PROFIBUS PA, FOUNDATION Fieldbus
|
Power supply
|
DC 12 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4-20 mA HART, 4-20 mA)
DC 12 to 35 V (4-20 mA HART, pulse/frequency/switch output, 4-20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
|
Hazardous area approvals
|
ATEX, IECEx, cCSAus
|
Other approvals and certificates
|
3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025),NAMUR, functional safety
PED, CRN, AD 2000
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614-1, similar to ASME IX (on request)
|
|
Steam
|
Measuring Principle
|
Vortex
|
Product headline
|
The flowmeter for accurate steam flow measurement with wet steam detection, available as compact or remote version. Integrated temp. measuring for mass/ energy flow of saturated steam. Suitable for a wide range of applications; optimized for steam applications.
|
Sensor features
|
Highest process safety – dualsens version enables redundant measurement. High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer. No maintenance – lifetime calibration. Wet steam detection and measurement for DN 25 to 100 (1 to 4"). Inlet run compensation.
|
Transmitter features
|
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology™. Display module with data transfer function. Robust two-chamber housing.
|
Nominal diameter range
|
DN 15 to 300 (½ to 12")
|
Wetted materials
|
Measuring tube: 1.4408 (C3FM); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4435 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Connection: 1.4404 (F316/F316L); CX2MW similar to Alloy C22, 2.4602; 1.4408 (CF3M)
|
Measured variables
|
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
|
Max. measured error
|
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (liquid): ±0.85%
Mass flow (steam,, gas): ±1.7 %
|
Measuring range
|
Liquid: 0.16 to 2412 m³/h (0.09 to 1420 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 2 to 32 166 m³/h (1.18 to 18 932 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
|
Max. process pressure
|
PN 40, Class 300, 20K
|
Medium temperature range
|
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)
|
Ambient temperature range
|
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
|
Sensor housing material
|
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
|
Transmitter housing material
|
AlSi10Mg, coated; 1.4404 (316L)
|
Degree of protection
|
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
|
Display/Operation
|
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
|
Outputs
|
4-20 mA HART (passive)
4-20 mA (passive)
Pulse/frequency/switch output (passive)
|
Inputs
|
4-20 mA (passive)
|
Digital communication
|
HART, PROFIBUS PA, FOUNDATION Fieldbus
|
Power supply
|
DC 12 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4-20 mA HART, 4-20 mA)
DC 12 to 35 V (4-20 mA HART, pulse/frequency/switch output, 4-20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
|
Hazardous area approvals
|
ATEX, IECEx, cCSAus
|
Other approvals and certificates
|
3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
PED, CRN, AD 2000
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614-1, similar to ASME IX (on request)
|
|