  • F 200 Vortex flowmeter

F 200 Vortex flowmeter

  • Product Item : 7F2C50-AABCCAAAABAD2SKA1
  • Category: Analysis
  • Measuring Principle:Vortex
  • Nominal diameter range:DN 15 to 300 (½ to 12")
  • Max. process pressure:PN 40, Class 300, 20K
  • Transmitter housing material:AlSi10Mg, coated; 1.4404 (316L)
  • INQUIRY   EMAIL

Proline Prowirl F 200 Vortex flowmeter

The flowmeter with detection of wet steam conditions, available as compact or remote version


The Prowirl F measuring tube is the first choice in heavy duty applications. The proven and patented capacitive DSC sensor ensures high precision measured values even under the toughest process conditions. Prowirl F 200 offers wetsteam detection and industry-compliant two-wire technology for seamless integration into existing infrastructures and control systems, as well as high operational safety in hazardous areas thanks to an intrinsically safe design, and a familiar installation procedure.

Benefits

  • Integrated temp. measuring for mass/ energy flow of saturated steam (opt.)

  • Highest process safety – dualsens version enables redundant measurement

  • High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer

  • No maintenance – lifetime calibration

  • Convenient device wiring – separate connection compartment

  • Safe operation – no need to open the device due to display with touch control, background lighting

  • Integrated verification – Heartbeat Technology™

Field of application

  • Preferred measuring principle for wet/saturated/superheated steam, gases & liquids (also cryogenic).

Device properties

  • Wet steam detection for DN 25 to 100 (1 to 4")

  • Inlet run compensation

  • Installation length according to industry standard

  • Display module with data transfer function

  • Robust two-chamber housing

  • Plant safety: worldwide approvals (SIL, Haz. area)

Features and specifications

Gas
Measuring Principle Vortex
Product headline The flowmeter for accurate steam flow measurement with wet steam detection, available as compact or remote version. Integrated temp. measuring for mass/ energy flow of saturated steam. Suitable for a wide range of applications; optimized for steam applications.
Sensor features Highest process safety – dualsens version enables redundant measurement. High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer. No maintenance – lifetime calibration. Wet steam detection and measurement for DN 25 to 100 (1 to 4"). Inlet run compensation.
Transmitter features Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology™. Display module with data transfer function. Robust two-chamber housing.
Nominal diameter range DN 15 to 300 (½ to 12")
Wetted materials Measuring tube: 1.4408 (C3FM); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4435 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Connection: 1.4404 (F316/F316L); CX2MW similar to Alloy C22, 2.4602; 1.4408 (CF3M)
Measured variables Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
Max. measured error Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (liquid): ±0.85%
Mass flow (steam, gas): ±1.7 %
Measuring range Liquid: 0.16 to 2412 m³/h (0.09 to 1420 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 2 to 32 166 m³/h (1.18 to 18 932 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
Max. process pressure PN 40, Class 300, 20K
Medium temperature range Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)
Ambient temperature range Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
Sensor housing material Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
Transmitter housing material AlSi10Mg, coated; 1.4404 (316L)
Degree of protection Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
Display/Operation 4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
Outputs 4-20 mA HART (passive)
4-20 mA (passive)
Pulse/frequency/switch output (passive)
Inputs 4-20 mA (passive)
Digital communication HART, PROFIBUS PA, FOUNDATION Fieldbus
Power supply DC 12 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4-20 mA HART, 4-20 mA)
DC 12 to 35 V (4-20 mA HART, pulse/frequency/switch output, 4-20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
Hazardous area approvals ATEX, IECEx, cCSAus
Other approvals and certificates 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, functional safety
PED, CRN, AD 2000
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614-1, similar to ASME IX (on request)
Liquids
Measuring Principle Vortex
Product headline The flowmeter for accurate steam flow measurement with wet steam detection, available as compact or remote version. Integrated temp. measuring for mass/ energy flow of saturated steam. Suitable for a wide range of applications; optimized for steam applications.
Sensor features Highest process safety – dualsens version enables redundant measurement. High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer. No maintenance – lifetime calibration. Wet steam detection and measurement for DN 25 to 100 (1 to 4"). Inlet run compensation.
Transmitter features Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology™. Display module with data transfer function. Robust two-chamber housing.
Nominal diameter range DN 15 to 300 (½ to 12")
Wetted materials Measuring tube: 1.4408 (C3FM); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4435 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Connection: 1.4404 (F316/F316L); CX2MW similar to Alloy C22, 2.4602; 1.4408 (CF3M)
Measured variables Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
Max. measured error Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (liquid): ±0.85%
Mass flow (steam, gas): ±1.7 %
Measuring range Liquid: 0.16 to 2412 m³/h (0.09 to 1420 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 2 to 32 166 m³/h (1.18 to 18 932 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
Max. process pressure PN 40, Class 300, 20K
Medium temperature range Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)
Ambient temperature range Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
Sensor housing material Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
Transmitter housing material AlSi10Mg, coated; 1.4404 (316L)
Degree of protection Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
Display/Operation 4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
Outputs 4-20 mA HART (passive)
4-20 mA (passive)
Pulse/frequency/switch output (passive)
Inputs 4-20 mA (passive)
Digital communication HART, PROFIBUS PA, FOUNDATION Fieldbus
Power supply DC 12 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4-20 mA HART, 4-20 mA)
DC 12 to 35 V (4-20 mA HART, pulse/frequency/switch output, 4-20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
Hazardous area approvals ATEX, IECEx, cCSAus
Other approvals and certificates 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025),NAMUR, functional safety
PED, CRN, AD 2000
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614-1, similar to ASME IX (on request)
Steam
Measuring Principle Vortex
Product headline The flowmeter for accurate steam flow measurement with wet steam detection, available as compact or remote version. Integrated temp. measuring for mass/ energy flow of saturated steam. Suitable for a wide range of applications; optimized for steam applications.
Sensor features Highest process safety – dualsens version enables redundant measurement. High availability – proven robustness, resistance to vibrations, temperature shocks & water hammer. No maintenance – lifetime calibration. Wet steam detection and measurement for DN 25 to 100 (1 to 4"). Inlet run compensation.
Transmitter features Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology™. Display module with data transfer function. Robust two-chamber housing.
Nominal diameter range DN 15 to 300 (½ to 12")
Wetted materials Measuring tube: 1.4408 (C3FM); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4435 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Connection: 1.4404 (F316/F316L); CX2MW similar to Alloy C22, 2.4602; 1.4408 (CF3M)
Measured variables Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
Max. measured error Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (liquid): ±0.85%
Mass flow (steam,, gas): ±1.7 %
Measuring range Liquid: 0.16 to 2412 m³/h (0.09 to 1420 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 2 to 32 166 m³/h (1.18 to 18 932 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)
Max. process pressure PN 40, Class 300, 20K
Medium temperature range Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F)
High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)
Ambient temperature range Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F)
Sensor housing material Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
Transmitter housing material AlSi10Mg, coated; 1.4404 (316L)
Degree of protection Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure
Display/Operation 4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available
Outputs 4-20 mA HART (passive)
4-20 mA (passive)
Pulse/frequency/switch output (passive)
Inputs 4-20 mA (passive)
Digital communication HART, PROFIBUS PA, FOUNDATION Fieldbus
Power supply DC 12 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4-20 mA HART, 4-20 mA)
DC 12 to 35 V (4-20 mA HART, pulse/frequency/switch output, 4-20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)
Hazardous area approvals ATEX, IECEx, cCSAus
Other approvals and certificates 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
PED, CRN, AD 2000
NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614-1, similar to ASME IX (on request)