Home      Products     Endress+Hauser     Analysis       CPS12D Digital ORP sensor
  • CPS12D Digital ORP sensor

CPS12D Digital ORP sensor

  • Product Item : CPS12D-7PA21
  • Category: Analysis
  • Measuring Principle:Sensor ORP / Redox
  • Application:Standard applications in process and environment. Long-term monitoring, processes with little changes.
  • Measurement range:-1500 mV - +1500 mV
  • Material:Glass electrode with PTFE diaphragm and gold pin or platinum ring.
  • INQUIRY   EMAIL

ORP sensor Orbisint CPS12(D)

Glass electrode for standard applications


Orbisint CPS12 is the all-rounder for ORP measurement in process and environmental applications. It performs reliably and accurately even in hazardous areas. Designed for low maintenance and a long operating life, the sensor offers you best value for money.
 

Benefits

  • Robust electrode with long poison diffusion path, requires low maintenance due to large, dirt-repellent PTFE ring diaphragm

  • Suitable for harsh applications: Process glass for highly alkaline media and pressure-stable up to 16 bar (232 psi)

  • Various measuring elements for application in oxidizing or reducing media

Field of application

  • Long-term monitoring and limit monitoring of processes with stable process conditions:
    - Chemical industry
    - Paper industry
    - Power plants (e.g. flue gas washers)
    - Incineration plants

  • Water treatment:
    - Drinking water
    - Cooling water
    - Well water

With ATEX, FM, CSA approval for use in hazardous areas.

Features and specifications

ORP / Redox
Measuring Principle Sensor ORP / Redox
Application Standard applications in process and environment. Long-term monitoring, processes with little changes.
Characteristic Very robust, heavy-duty / high pressure, integrated potential matching.
Measurement range -1500 mV - +1500 mV
Measuring principle - Gel compact electrode with teflon diaphragm and double gel reference - measuring part as gold-pin or platinum-ring.
Design - All shaft lengths with temperature sensor - advanced gel technology
Material Glass electrode with PTFE diaphragm and gold pin or platinum ring.
Dimension Diameter: 12 mm (0.46 inch)
Shaft lengths: 120, 225, 360 and 425 mm
(4.68, 8.77, 14.04 and 16,57 inch)
Process temperature max. 135°C (274°F)
Process pressure max. 16bar (230psi)
Ex certification Simple operational device, no certification necessary.
Connection TOP68 connection head
Ingres protection IP68
Additional Certifications Biocompatibility acc. ISO 10993-5 and USP (current version).
TÜV certificate for pressure resistance with min. triple overpressure safety.