Home      Products     Endress+Hauser     Analysis       CLS54D Digital conductivity sensor
  • CLS54D Digital conductivity sensor
  • CLS54D Digital conductivity sensor
CLS54D Digital conductivity sensorCLS54D Digital conductivity sensor

CLS54D Digital conductivity sensor

  • Product Item : CLS54D-AACS1A4A
  • Category: Analysis
  • Measuring Principle:Inductive
  • Application:Food, beverage and pharmaceutical industries and in biotechnology.
  • Measurement range:100 µS/cm - 2000 mS/cm Cell constant k: 6,3 1/cm
  • Design:Hygienic design with food-grade virgin PEEK body in jointless and creviceless design including all hygienic certifications.
  • INQUIRY   EMAIL

Digital conductivity sensor Indumax CLS54D

Memosens toroidal conductivity sensor for hygienic and sterile applications


The Indumax CLS54D inductive conductivity sensor protects all of your products and processes that must meet the strictest hygienic and sterile demands. Its outstanding performance combined with a unique hygienic design ensures highest safety and quality of your products. Thanks to Memosens digital technology, Indumax CLS54Dcombines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration, and facilitates predictive maintenance.

Benefits

  • Unique hygienic design avoids contamination

  • With all hygienic certificates required in hygienic and sterile applications, compliant with EG 2023/2006 and 1935/2004

  • Food grade virgin PEEK body without joints nor crevices

  • Certified biocompatibility according to USP class VI

  • Suitable for cleaning-in-place (CIP) and sterilization-in-place (SIP)

  • Convenient maintenance strategy with Memobase Plus for easy data and sensor management

  • Reliable values thanks to active connection monitoring and EMC protection

Field of application

Indumax CLS54D measures toroidal conductivity in hygienic food & beverage and life science applications, such as:

  • Phase separation of product/water and product/product mixtures in pipe systems

  • Control of CIP (cleaning in place) processes in the return channel

  • Concentration control in remaking of CIP cleaning agents

  • Product monitoring in pipe systems, bottling plants, quality assurance

  • Leakage monitoring

Its unique hygienic design is certified by EHEDG and 3-A.

Features and specifications

Thermometer  
Measuring Principle Inductive
Application Food, beverage and pharmaceutical industries and in biotechnology.
Characteristic Specifically intended for use in hygienic applications and especially suitable for following applications:
Phase separation, control of CIP, concentration control, product monitoring, leakage monitoring.
Measurement range 100 µS/cm - 2000 mS/cm Cell constant k: 6,3 1/cm
Measuring principle Inductive conductivity measurement
Design Hygienic design with food-grade virgin PEEK body in jointless and creviceless design including all hygienic certifications.
Material In contact with medium: Virgin PEEK
Surface roughness: Ra<0.8µm
Process temperature Process temperature: -10°C - +125°C
(14°F - 257°F)
Sterilisation: 150°C/5 bar (max. 60 min.)
(302°F / 72.5 psi)
Process pressure 12 bar(174 psi)/20°C(68°F)
8 bar(116 psi)/125°C(257°F)
temperature sensor Pt1000 (class A)with a temperature response time T90 < 27s.
Connection Dairy pipe, Aseptic-fitting, Clamp ISO 2852, SMS-2", Varivent N DN 40-125, Neumo Biocontrol D50.
Ingres protection IP68/NEMA6P
Additional Certifications FDA, EHEDG, 3-A,Bioreaktivität (USP class VI part 87 and 88).