Digital free chlorine sensor
Memosens CCS51D
Memosens sensor for drinking water, pool and process water as well as utilities in all industries
|
Field of application
Memosens CCS51D is a robust, low-maintenance sensor for free chlorine. It provides high-precision measurement with long-term stability in drinking water, pool and process water and ensures consistent disinfection monitoring even at low water volumes. The free chlorine sensor features extremely fast response times guaranteeing efficient process control. Memosens CCS51D is not the latest Memosens generation. More information on the new Memosens CCS51E sensor with extended functionality here.
Memosens CCS51D is a free chlorine sensor for skid builders and end customers. It measures reliably in:
-
Drinking water - to ensure reliable disinfection
-
Process water - for hygienic and safe packaging and bottling of products
-
Utilities - to detect the absence or presence of free chlorine
-
Pool water - to dose disinfectant efficiently
|
Benefits
-
The right sensor version for every application: From trace measurement up to free chlorine concentrations of 200 mg/l.
-
Fast response time (t90 < 25 s) provides an accurate process view and enables prompt reaction to process changes as well as efficient process control.
-
Increased process safety: precise and long-term stable measurement ensures consistent process monitoring and allows for individually adapted disinfectant dosing.
-
The low-maintenance, amperometric sensor reduces the cost of ownership of the measuring point, particularly compared to colorimetric measuring systems.
-
More process up-time thanks to fast sensor exchange: pre-calibrate the sensor in your lab and then swap it into your process with plug & play.
-
Connection to the Liquiline multiparameter transmitter allows for easy combination with other relevant parameters of liquid analysis such as pH and ORP.
|
-
-
Application
Ensuring reliable disinfection in drinking water
Process water
Dosing disinfectant efficiently in pool water
Detect the absence or presence of free chlorine in Utilities
Guarantee food safety and provide hygienic packaging and bottling
-
Characteristic
Amperometric measurement of dissolved free chlorine
-
Measurement range
Trace: 0 to 5 mg/l HOCl
Standard: 0 to 20 mg/l HOCl
High: 0 to 200 mg/l HOCl
-
Measuring method
Closed, membrane covered measuring cell
Reduction of free chlorine at the cathode
-
Design
Closed amperometric 2-electrode measuring cell with PVDF membrane
-
-
Material
Sensor shaft: POM
Membrane: PVDF
Membrane cap: PVDF
Sealing ring: FKM
-
Dimension
Diameter: 25 mm (0.98 inch)
Length: 161 mm (6.34 inch)
-
Process temperature
0 to 55 °C (32 to 130 °F), non-freezing
-
Process pressure
Max. 1 bar (max. 14.5 psi)
-
Temperature sensor
10k NTC integrated (Memosens)
-
Connection
Inductive, digital connection head with Memosens
CCS51D-AA11AD
-
Approval:AA
Non-hazardous area + UKCA marking
Application:11
Process water, drinking water
Measuring Range:AD
5 mg/l
-
|